The Renishaw XL-80 laser interferometer offers the ultimate in high performance measurement and calibration for motion systems, including CMMs and machine tools. Fast, accurate and extremely portable, it is no surprise the XL-80 system is the best selling calibration laser. Together with advanced software solutions and superior performance, the XL-80 can significantly improve your business performance.
The Renishaw XL-80 laser produces an extremely stable laser beam, with a wavelength that is traceable back to national and international standards.
- Accuracy and dynamic performance
- Ease of use
- Portability
- Laser frequency accuracy ±0.05 ppm (parts per million) over 3 years is achieved by thermal control of the laser tube's length to within nanometres.
- Linear measurement accuracy ±0.5 ppm over the whole environmental range ie, from 0 ºC - 40 ºC (32 ºF - 104 ºF) and 650 mbar - 1150 mbar. Readings can be taken at up to 50 kHz, with a maximum linear measurement speed of 4 m/s and a linear resolution of 1 nm; even at maximum speed.Integrated USB means there is no requirement for a separate laser-to-PC interface.
- LED status lights indicate the laser status and signal strength, providing a convenient alternative to the software's “on-screen” indicators.
- Switchable between standard (40 m) and long (80 m) range modes.
- Analogue I/O port allows for analogue signal output and a trigger signal input.
- Warm-up time is less than 6 minutes.
- External, switch mode power supply ensures 90 V - 264 V flexibility in input voltage.
- CARTO software suite allows measurement of linear, angular, rotary axis, straightness and squareness. For other measurement modes LaserXL is required.
|